68
OPERATION
HOW TO ATTACH AND SECURE THE RFC
1. Rotate the lever to its “OFF” position,
2. Position the mobile flange against the fixed flange using the alignment notches, then rotate the
lever in its “ON” position,
3. Apply a gentle pressure on the mobile flange so as to make sure that it is firmly in position (or does
not wobble),
4. Engage the Lock screw over 3 or 4 turns inside the fixed flange. The screw must prevent the mobile
from being taken off, while not been completely tightened. Tightening this screw too firmly can
degrade the coupling injection ratio.
HOW TO DETACH THE RFC
1. Loosen the Lock screw,
2. Move the lever in its “OFF” position,
3. Release the mobile flange,
4. Insert the cap cover inside the aperture of the mobile flange
DANGER:
LASER RADIATION
Make sure to turn off the laser emission any time the RFC is being manipulated.