COMMAND AND QUERIES FOR THE LBX LASERBOXX
Purpose
Query
(and
aliases)
Result and syntax
Retrieves the regulation loop. See also
“?APC”.
?ACC
1” when the ACC mode is set
0” when the ACC mode is not set
Retrieves the status for analog modulation
?AM
?EXT
1when the external modulation is engaged
0” when the set point is internal
Retrieves the regulation loop. See also
“?ACC”.
?APC
1” when the APC mode is set
0” when the APC mode is not set
Retrieves the temperature of the base
plate of the laser head
?BT
22.1” for 22.1°C
Purpose
Command
(and
aliases)
Syntax and examples
Modifies the power set point.
Only for models that accept power
adjustment OPT-PWR-LCX
IP
P
IP 100 (IP,<space>,100) sets the optical power at 100%
of the nominal power.
IP 50 (IP,<space>,50) sets the optical power at 50% of
the nominal power.
P 250.3 (P,<space>,250,<dot>,3) sets the optical power
at 250.3mW
Switches the temperature regulation loop
ON or OFF. This regulation loop is
necessary for the emission to occur.
Switching it off helps reducing the power
consumption.
T
T 0 (T,<space>,0): switches the emission off
T 1 (T,<space>,1): switches the emission on