91
1 Emission is off
2 Emission is on at low power
Retrieves the laser’s measured output
power (in milliwatts)
?P
IP?
53.6
“53.6 mW”
Queries the state of the laser module
(see section “Troubleshooting” for more
information concerning the laser status)
?STA
DL?
?STA returns a number corresponding to one of the following
status:
1 : Warm-up phase
2 : Warm-up phase completed, emission OFF
3 : Emission ON
5 : Alarm present
6 : Sleep mode
7 : Searching for SLM point
“DL?” returns one of the following text messages:
Laser off
Setting temperature
Waiting for stabilization
Starting up
Laser on
Laser alarm
Retrieves the status of the temperature
regulation loop
?T
T?
0 Temperature regulation loop is deactivated and the
emission is off
1 Temperature regulation loop is activated, the emission
is on