Retrieves the temperature of the base
plate of the laser head (expressed
in Celsius degrees)
Retrieves the cause of the latest alarm
(see section “Troubleshooting” for more
information about alarms)
“?F” returns a number corresponding to one of the following
alarms:
“0” : No alarm
“1” : Diode current
“2” : Laser power
“3” : Power supply
“4” : Module temperature
“5” : Base temperature
“7” : Interlock
“8” : User-generated alarm (using the command RST)
AL? returns one of the following messages:
“Laser Head Temperature”
“Interlock”
“Power Supply”
“Other”
“No alarm”
Retrieves the laser operation time, in hours
“49”
“TIME= 49 hrs, 38 min”
Retrieves the unit’s serial number
“LAS-XXXXX”, where XXXXX is a five-digit number
Retrieves the type of laser
“LCX-532-50” for a 50mW LCX emitting at 532nm
Retrieves the emission status