89
Sets the power set point on source number
n (where n is a number from 1 to 6). On
sources linked to a modulator (AOM, MPA),
the set point of that modulator will be
affected. The argument is an absolute
optical power.
PLn
PL1 10.1 (PL1,<space>,10.1) sets the power of laser
source L1 at 10.1mW
Retrieves the power set point on source
number n (where n is a number from 1 to
6). On sources linked to a modulator (AOM,
MPA), the set point of that modulator will
be returned.
?PLn
40.0” for a set point of 40.0 mW
Sets the power set point on source number
n (where n is a number from 1 to 6). On
sources linked to a modulator (AOM, MPA),
sets the set point of that modulator. The
argument is a percentage of the nominal
power.
PPLn
PPL2 25.2 (PL2,<space>,25.2) sets the power of laser
source L3 at 25.2% of its nominal power
Retrieves the power set point on source
number n (where n is a number from 1 to
6), as a percentage of the nominal power.
On sources linked to a modulator (AOM,
MPA), the set point of that modulator will
be returned.
?PPLn
21.0” for a set point of 21.0 %
Open or close shutter on channel n (where
is a number from 1 to 4)
SHn
SH1 1 (SH1,<space>,1): opens the shutter on channel 1
SH1 0 (SH1,<space>,0): closes the shutter on channel 1
SENDING COMMANDS TO THE INDIVIDUAL LASER SOURCES
It is possible to address to the individual LaserBoxx sources by sending command or queries to the combiner
using the prefix “Ln”, where n is the number of the source to be addressed (L1 to L6).
COMMAND AND QUERIES FOR THE LCX LASERBOXX
Purpose
Query
(and
aliases)
Result and syntax