88
1 to 6). Refer to the specific commands for
LBX and LCX in the following tables.
L3 INF? (L3,<space>,INF?): Get the model name of
source number 3
Retrieves the status of the combiner
?ST
Lasers off” when the actuated key master control is
off
Lasers on” when the actuated key master control is on
and all the sources in use are emitting
Starting-up” when at least on laser sources in use is in
“starting-up” status
Alarm” when at least one of the sources is in “alarm”
status
Stand-by” when all laser sources are in “stand-by”
status
Blanking” when the blanking function is activated
Sets the bias voltage v on the AOM n
(where n equals 1 or 2, and v is between 0
and 50). This value determines the
extinction contrast of the AOM. We
recommend a value of 20 for v.
TA n v
TA 1 20 (TA,<space>,1,<space>,20): sets the bias
voltage to 20 on the AOM 1.
Retrieves the value of the bias voltage on
AOM n (where n equals 1 or 2).
?TA n
20
Power and operational settings:
Purpose
Query
Syntax and examples
Sets the position of the flip mirror number
p (where p is a number from 1 to 5). Each
flip mirror can be set to either “0” or “1”
positions
FMn
FM1 1 (FM1,<space>,1) toggle the flip mirror number 1 to
its position “1”
Sets the position of the fast mirror number
p (where p is a number from 1 to 2). Each
flip mirror can be set to either “0” or “1”
positions
GAn
GA1 1 (GA1,<space>,1) toggle the fast mirror number 1 to
its position “1”
Sets the transmission ration of the “M-
Dual” module
IPA
IPA 40 (IPA,<space>,40) sets a transmission ratio of 40%
on channel 1