95
Sets the laser power, then saves this value
as the new default.
P
P 20 (P,<space>,20) sets the optical power at 20mW.
Values allowed between 0 and the maximum power (see
the query ?MAXLP).
Sets the laser power, but does not record
this value in memory, thus sparing stress on
the internal EEPROM.
Prefer this command to “P” if you need to
change the power at high frequency (ten
times per second or higher).
PM
PM 20 (PM,<space>,20) sets the optical power at 20mW.
Values allowed between 0 and the maximum power (see
the query ?MAXLP).
Switches the temperature regulation loop
ON or OFF. This regulation loop is
necessary for the emission to occur.
Switching it off reduces the power
consumption.
T
T 0 (T,<space>,0): switches the emission off, and sets the
operating status to “sleep” mode (see “STA”).
T 1 (T,<space>,1): switches the emission back on. The
unit is no longer in “sleep” mode.