41
ACU-TROL
®
AK110 Chemical Controller Installation and User's Guide
AKColor PPM Troubleshooting
pH Troubleshooting
Problem Solution
PPM Reading is 0 or 9.99 1. Check Flow to AKColor Flow Chamber
2. Check Chemical Levels
3. Check Discharge from DPD and
Buffer Solenoids
4. Clear Calibrations
Chemical Reagents are being 1. Check sample rate, decrease to 5 min or used too
quickly greater.
2. Check the DPD and Buffer solenoids for leaks
Measured PPM Level is lower 1. Check sample flow solenoid for leaks
than the hand check reading 2. Recalibrate the PPM reading with pH at
setpoint.
Theclearvoltageisnot3.5 1.AdjustthepotentiometerintheAKColor
Problem Solution
pH not reading correctly 1. Use the test strip to verify pH
2. Confirm that the BNC sensor wires are
installed correctly
3. Install a new BNC sensor wire.
4. Check the manufacture date of the sensor
5. Confirm that the flow rate in the AK1200 Flow Cell is between
0.5 and 1.5 liters per minute.
pH readings response slow to
to changes in the water 1. Confirm that there is flow in the Flow Cell
2. Clean the pH sensor using the Acu-Trol pH
sensor or cleaning procedure listed in this manual.
pH readings inaccurate 1. Recalibrate the pH reading