Technical Data
Details of the Directives:
ATEX Directive 2014/34/EU
Ignition protection type: Gas ATEX category 3G Ex nA IIC T4 Gc X
Dust ATEX category 3D Ex tc IIIC T135°C Dc X
FM - Factory Mutual
NI/I/2/ABCD/T5; +5°C < Ta < 55°C IP64
(cables and cable glands are not part of the FM approval of the device and are therefore
not fitted at the factory.)
c UL us - Underwriters Laboratories (Canada and USA)
UL 61010-1 AND CSA C22.2 NO. 61010-1
Restrictions: Application area: 0 to +55°C,
Indoor use,
power supply with class-2 power supply unit
6.4. Information on the additional plates
Additional plates indicate additional approvals and special operating conditions.
Warning sign for use of the device in potentially explosive atmosphere
Line 1
Line 2
Line 3
Line 4
Line 1 Specifications according to ATEX directive (gas) / ambient temperature
Line 2 Specifications according to ATEX directive (dust) / degree of protection specification
Line 3 WARNING POTENTIAL ELECTROSTATIC CHARGING
Line 4 HAZARD SEE INSTRUCTIONS
Additional plate for devices with UL approval
UL label with UL file no.
Note on use of power supply unit according to
NEC Class 2
Permissible supply voltage (24 V DC)