Installation
18
Make sure that the detector is off (circuit breaker at 0, the control panel screen is
off) and in a horizontal position.
Open the cover.
Remove the oil fill cap (1) from the rotary vane pump (2).
Put the funnel (included with the detector) in the oil fill opening (3).
Fill with oil to the highest level.
Replace the oil fill cap tightly (1).
5.5 Connecting the purge circuit
The Dry Model's purge system can be connected to a purge circuit and can be open/
closed depending on the settings.
The Wet Model's purge system is always closed and cannot be set by the operator.
To guarantee best performance, the nitrogen supply must be clean and filtered, with the
following characteristics:
– relative excess pressure : 200 hPa
– flow rate: 50 sccm
Attach the nitrogen pipe to the connector (see 4.2).
Regulate the purge flow rate.
Max
Min
3
NOTICE
Tracer gas concentration
When the air purge is used, we recommend performing the leak detection in a ventilated
room.
WARNING
Pressurised circuit
To work safely on the product, the operator must:
Install a manual valve on the nitrogen circuit at a distance of 3 m from the product, so
that the nitrogen supply can be locked.