Advanced settings
36
Screen settings Brightness High / Low High
Contrast 0 - 100 50
Panel Off None /
15 min /
30 min /
1 h /
2 h /
4 h
None
Paging Function Without RC 500 WL remote control detected - None
With RC 500 WL remote control detected Off / On Off
Reset panel parame-
ters
Function launching - -
Access / Password Password 0000 - 9999 5555
Set Points Menu Acess Lock /
Unlock
Unlock
Test Menu Access Lock /
Unlock
Unlock
Spectro Menu Access Lock /
Unlock
Unlock
Maintenance Menu Access Lock /
Unlock
Unlock
Configuration Menu Access Lock /
Unlock
Unlock
Advanced Menu Access Lock /
Unlock
Unlock
User Level Restricted /
Medium Access /
Full Access
Full Access
Change Password 0000 - 9999
CONFIGURATION
Selection Choice - Setting
limit
Initial settings
ADVANCED
Selection Choice -
Setting limit
Initial settings
Leal Detection Start Up Timer 0 - 1 h 10 s
Background Suppres-
sion
Activation Off / On On
Crossover Pressures Gross Leak 2.5·10
+1
- 1·10
+1
2.5·10
+1
Normal 5·10
-1
- 1·10
-1
5·10
-1
Calibration Calibration Operator /
Start-Up /
Manual
Operator
Calibration
checking
Checking Operator /
Automatic
Operator
Frequency (If
automatic)
Cycles 0 - 9999 50
Hours 0 - 9999 10
Analyzer Cell Filament Selected 1 / 2 1
Filament Off /On On
Triode Pression - -
Electric Zero - -
CalibrationValve - Closed
Target Value - -
Acceleration Voltage (V) - -
Emission (mA) - -
Sensitivity Coefficient - -
Internal Temperature (°C) - -