Draft - 1 Aug 08
Telemetry Functions at the Information Center
Telemetry Functions at the Information Center & TeleMon
9-11
Suspend
Duration
Sets the alarm suspend duration
time for each assigned device on
the Information Center.
1, 2, or 3 minutes 2 minutes
Battery
Gauge on
Information
Center
Display/disable a battery gauge
for each assigned device on the
Information Center.
Note—Set to disable if
rechargeable batteries are being
used.
enable
disable
enable
(battery
gauge is
displayed)
RF Auto
Shutoff
Enable/disable RF operation
during an extended situation of
all leads off for more than 10
minutes and the SpO
2
sensor is
not connected.
enable
disable
enable
Fixed Pacer
Amplitude
Sets the appearance of the pacer
spikes to a fixed size as they
appear in the patient window.
enable
disable
disable
Autopair Enable/disable the autopairing of
the transceiver and the
IntelliVue Patient Monitor at the
Information Center.
enable
disable
enable
Enable SpO
2
Enable/disable the SpO2
measurement at the Information
Center or Telemon.
enable
disable
enable
Unit Settings - Telemetry Setup
Control Function Settings
Factory
Default

Table of Contents