Draft - 1 Aug 08
Telemetry Functions at the Information Center
Telemetry Functions at the Information Center & TeleMon
9-17
Systolic Low Increment/decrement NBP low
alarm limit by 1.
Limit Maximum is 155
Limit Minimum is 28 (Adult)
Limit Maximum is 255
Limit Minimum is 28 (Ped.)
90 Adult
70 Pediatric
Diastolic
High
Increment/decrement NBP high
alarm limit by 1.
Limit Maximum is 260
Limit Minimum is 90 (Adult)
Limit Maximum is 260
Limit Minimum is 45 (Ped.)
90 Adult
70 Pediatric
Diastolic Low Increment/decrement NBP high
alarm limit by 1.
Limit Maximum is 85
Limit Minimum is 28 (Adult)
Limit Maximum is 40
Limit Minimum is 28 (Ped.)
50 Adult
40 Pediatric
Mean High Increment/decrement NBP high
alarm limit by 1.
Limit Maximum is 260
Limit Minimum is 65 (Adult)
Limit Maximum is 260
Limit Minimum is 55 (Ped.)
110 Adult
90 Pediatric
Mean Low Increment/decrement NBP high
alarm limit by 1.
Limit Maximum is 60
Limit Minimum is 28 (Adult)
Limit Maximum is 50
Limit Minimum is 28 (Ped.)
60 Adult
50 Pediatric
Unit Settings - NBP Setup
Control Function Settings
Factory
Default

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals