Telemetry Functions at the Information Center
9-16 Telemetry Functions at the Information Center & TeleMon
6-wire, ECG4 Set the unit default lead. I, II, III, MCL, AVR, AVL,
AVF, V
1
, V
2
, V
3
, V
4
, V
5
, V
6
,
V
7
, V
8
, V
9
, V
3R
, V
4R
, V
5R
V
5
; V lead
choice is
determined
by Va and
Vb settings
Unit Settings - Default Leads
Control Function Settings
Factory
Default
Unit Settings - NBP Setup
Control Function Settings
Factory
Default
Patient Type Set patient type used for NBP
alarm limits.
Adult
Pediatric
Adult
NBP Alarm Set NBP alarm notification. Systolic or Diastolic
Systolic
Diastolic
Mean
Off
Systolic or
Diastolic
Systolic High Increment/decrement NBP high
alarm limit by 1.
Limit Maximum is 260
Limit Minimum is 160 (Adult)
Limit Maximum is 260
Limit Minimum is 75 (Ped.)
160 Adult
120 Pediatric