Physiologic Alarms
5-8 Alarms
* NON-SUSTAIN VT Yellow A run of Vs having a ventricular Heart
Rate greater than V-Tach limit but lasting
for less than the V-Tach Run limit.
ST/AR
Enhanced
Arrhythmia
* PACER NOT CAPT Yellow No QRS for 1.75 x the average R-R
interval with Pace Pulse (paced patient
only).
ST/AR
Basic &
Enhanced
Arrhythmia
* PACER NOT PACE Yellow No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only).
ST/AR
Basic &
Enhanced
Arrhythmia
* PAIR PVCs Yellow Two consecutive PVCs between non-
PVCs.
ST/AR
Enhanced
Arrhythmia
* PAUSE Yellow No QRS detected for x seconds. Choices
of >1.5 to 2.5 seconds.
ST/AR
Enhanced
Arrhythmia
* PVCs >xxx/MIN Yellow PVCs within one minute exceed by the
PVCs/min limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
** QTc High Yellow QTc value has exceeded the QTc High
limit for more than 10 minutes
ST/AR
Arrhythmia
**dQTc High Yellow dQTc High value has exceeded the dQTc
High Limit for more than 10 minutes.
ST/AR
Arrhythmia
Physiologic (Patient) Alarms
Alarm Text Priority Condition Source