Draft - 1 Aug 08
Physiologic Alarms
Alarms
5-9
* R-ON-T PVCs Yellow For Heart Rate less than 100, a PVC with
R-R interval less than 1/3 the average
interval followed by a compensatory
pause of 1.25 x average R-R interval, or 2
such Vs without a compensatory pause
occurring within 5 minutes of each other.
(When Heart Rate is greater than 100, 1/3
R-R interval is too short for detection.)
ST/AR
Enhanced
Arrhythmia
* RUN PVCs Yellow Run of PVCs greater than or equal to 2. ST/AR
Enhanced
Arrhythmia
* SVT Yellow Run of SVPBs greater than or equal to
SVT Run limit and with SVT Heart Rate
greater than the SVT Heart Rate limit.
ST/AR
Enhanced
Arrhythmia
* VENT BIGEMINY Yellow A dominant rhythm of N, V, N, V (where
N= supraventricular beat, V=ventricular
beat).
ST/AR
Enhanced
Arrhythmia
* VENT RHYTHM Yellow A dominant rhythm of adjacent Vs greater
than Vent Rhythm limit and ventricular
Heart Rate less than V-Tach limit.
ST/AR
Enhanced
Arrhythmia
* VENT TRIGEMINY Yellow A dominant rhythm of N, N, V, N, N, V
(where N=supraventricular beat,
V=ventricular beat).
ST/AR
Enhanced
Arrhythmia
Physiologic (Patient) Alarms
Alarm Text Priority Condition Source

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals