Technical Alarms (INOPs)
5-12 Alarms
CENTRAL: TELE ONLY
Source - Patient Monitor
Hard System connectivity via the
transceiver is limited (no
alarms, only local
numerics) when in
companion mode and host
monitor does not have
system connectivity. Only
the transceiver’s
measurements and alarms
can be displayed at the
Information Center.
For more information only.
CHECK ECG SETTINGS
Source - Telemetry
Hard Synchronization of ECG
settings between the
monitor and the
Information Center has
failed.
Check that the ECG settings
in use are appropriate.
Note—When transitioning
between networked and non-
networked monitoring , this
INOP will display. Pressing
the Silence button will
dismiss the INOP.
CHECK ECG SOURCE
Source - Patient Monitor
Yellow
Technical
Alarm
The transceiver and the
monitor both have valid
ECG signals.
Unpair the transceiver and
monitor or remove the
leadset from the device that
is no longer the ECG source.
CHECK PAIRING
Source - Telemetry
Yellow
Technical
Alarm
There is a problem with
device pairing.
Check that the bedside
monitor and telemetry device
are correctly paired.
CHECK SpO
2
T Settings
Source - Telemetry
Hard Synchronization of SpO
2
T
settings between the
monitor and the
Information Center has
failed.
Check that the SpO
2
T
settings in use are
appropriate.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do