Technical Alarms (INOPs)
Alarms
5-19
TELE SYNC UNSUPP
Source - Telemetry
Hard Incompatible MMS The MMS in use does not
support synchronization of
ECG and SpO2 settings
between the monitor and
Information Center after a
transceiver has been paired.
The MMS must be revision
E.00 or above.
TELE UNSUPPORTED
Source - Patient Monitor
Hard The transceiver is not
supported for direct
connection to the monitor.
Make sure the transceiver is
the correct revision.
TELE WEAK SIGNAL
Source - Telemetry
Soft • Patient is at outer
range of the radio
coverage area.
• Telemetry pack is
receiving a weak
signal with high data
loss from the AP.
• Condition exists for
multiple devices in a
specific area
• Return patient to the
coverage area.
• If patient is in close
proximity to AP,
replace telemetry pack.
Contact service.
• The AP covering the
specific area is
suspect. Contact
Service.
TRANSMITTER MALF
Source - Telemetry
Hard Transceiver malfunction Replace and contact Service.
TRANSMITTER OFF
Source - Telemetry
Hard RF Auto Shutoff after 10
minutes of all leads off and
no SpO
2
sensor connected.
• Reattach ECG leads to
patient.
• Reattach SpO2 sensor.
• Press the Check
button.