Technical Alarms (INOPs)
Alarms
5-19
TELE SYNC UNSUPP
Source - Telemetry
Hard Incompatible MMS The MMS in use does not
support synchronization of
ECG and SpO2 settings
between the monitor and
Information Center after a
transceiver has been paired.
The MMS must be revision
E.00 or above.
TELE UNSUPPORTED
Source - Patient Monitor
Hard The transceiver is not
supported for direct
connection to the monitor.
Make sure the transceiver is
the correct revision.
TELE WEAK SIGNAL
Source - Telemetry
Soft Patient is at outer
range of the radio
coverage area.
Telemetry pack is
receiving a weak
signal with high data
loss from the AP.
Condition exists for
multiple devices in a
specific area
Return patient to the
coverage area.
If patient is in close
proximity to AP,
replace telemetry pack.
Contact service.
The AP covering the
specific area is
suspect. Contact
Service.
TRANSMITTER MALF
Source - Telemetry
Hard Transceiver malfunction Replace and contact Service.
TRANSMITTER OFF
Source - Telemetry
Hard RF Auto Shutoff after 10
minutes of all leads off and
no SpO
2
sensor connected.
Reattach ECG leads to
patient.
Reattach SpO2 sensor.
Press the Check
button.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals