Draft - 1 Aug 08
Technical Alarms (INOPs)
5-18 Alarms
TELE CONFIG
UNSUPPORTED
Source - Patient Monitor
Hard Transceiver not supported
(companion mode).
Make sure the transceiver is
the correct revision.
Contact Service.
TELE DISCONNECTED
Sourece - Telemetry
Red,
Yellow or
Hard
Technical
Alarm
The interface cable is
not connected.
Short-range radio
connection between
the transceiver and
the patient monitor
has been lost.
If the disconnection is not
intentional:
Re-connect the
interface cable.
Identify and remove
interference sources
such as bluetooth
devices, DECT
phones, microwaves,
etc.
Reduce the number of
devices equipped with
short-range radio
capability.
TELE INCOMPATIBLE
Source - Patient Monitor
Hard Transceivers equipped with
SRRA are not supported
with this revision of the
Information Center.
Check equipment
configuration/compatibility.
Contact Service.
TELEMETRY STANDBY Soft Information Center standby
mode timer is active, or
patient was not returned to
telemetry coverage area.
There is no data from bed.
Cancelled when patient is
removed from Standby.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals