3. Insert the suction tubes into the suction tube cleaning holder.
NOTE
Make sure that the suction tube cleaning holder itself is in place (i.e.
pushed to the bottom of the suction cleaning unit. See step 6 for a
picture of the unit).
NOTE
Make sure that a plug is inserted into each empty suction tube holder in
the suction tube cleaning holder to prevent the concentrate from spilling.
4. Lift the dosing cubic from the suction cleaning unit and fill it with
Planmeca approved suction disinfectant.
The dosing scale indicates the required amount for 1 or 2 suction tubes.
1 1 suction tube
2 2 suction tubes
5. Place the dosing cubic back to the suction cleaning unit.
6. The green indicator light of the start button is on when the cleaning
program can be started. Start the cleaning by pressing the green start
button.
The start button indicator light is flashing during the cleaning procedure
and the control panel informs you about the progress of the procedure.
The cleaning program can be interrupted by pressing the yellow cancel
button. H 43 is displayed.
After interrupting the cleaning program you can start it again by filling the
dosing cubic and pressing the green start button. The program will start
from the beginning.
23 Flushing and cleaning programs
116 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Classic v2 and is the answer not in the manual?

Planmeca Compact i Classic v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Classic v2
CategoryDental equipment
LanguageEnglish

Related product manuals