EasyManua.ls Logo

Planmeca Compact i Classic v2 - Instrument Related Error Messages

Planmeca Compact i Classic v2
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 42.1 Suction cleaning completed. Return suction
tubes to suction holder.
H 43 Suction cleaning cancelled. Return suction
tubes to suction holder.
H 44 Incorrect suction tube amount detected.
Return them to suction holder, place them
back to suction tube cleaning holder and
start cleaning.
H 45 Set WMS to Domestic Water Supply -mode
and wait until WMS bottle is full. Then start
suction cleaning.
Suction cleaning can not be used when
WMS is set to Bottled Water -mode.
H 46 Suction use is forbidden or there are no
suction tubes in dental unit configuration.
The value of service mode n.83 is 0.
H 47 Suction disinfectant pressure is low. (In
automatised suction cleaning: Fill suction
disinfectant container).
H 47.1 Suction disinfectant level low. Fill suction
disinfectant container.
H 47.2 Suction disinfectant container filled.
H 48 WMS must be set to Bottled water off and
Bottle on -modes before starting suction
cleaning.
Bottled water off = domestic water is used
Bottle on = bottle is used; domestic water is
run through the bottle
H 50 No cup in cup holder. Sensor does not
recognise cup when pressing cup fill button.
Place a cup on the cup holder. Or change
the setting of service mode n 115.
H 51 Bowl is above patient chair when driving
chair up.
If the unit has a bowl, move the bowl to the
rest position. If the unit has an OP delivery
arm safety switch, check that the OP
delivery arm is not blocking the chair
movement.
H 52 Patient chair can not be driven upward
because safety switch of side delivery arm
is released.
H 53 Patient chair can not be driven downward
because safety switch of side delivery arm
is released.
H 54 Cup is inserted in cup holder. Remove the cup and make sure the cup fill
tube is above the bowl.
H 55 Disinfectant container attached. To start
waterline cleaning, select instruments and
press Waterline cleaning button.
H 57 Vision action failed. Select patient in
Romexis.
Select patient in Planmeca Romexis before
performing Solanna Vision action.
H 60 WMS: Container almost empty in Domestic
water supply -mode.
Reduce water use for a moment.
H 61 WMS: Container almost empty in Bottled
water -mode.
Reduce water use and fill the container as
soon as possible.
27 Help and error messages
User's manual Planmeca Compact i 145

Table of Contents

Related product manuals