Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 62 WMS: Container switch is turned off. Switch on the container switch.
H 63 Cuspidor door is open. Close the door.
H 64 Fill container with disinfectant over upper
sensor.
H 65 Fill container with water over upper sensor.
H 66 Wait 8 hours or switch unit off for the night.
H 67 Instrument flush cancelled. Return
instruments to instrument console.
H 68 Romexis connection lost. Make sure the Ethernet cable is attached
and Romexis is running.
H 69 Unable to create a new user. Maximum
amount of local users exceeded.
H 70 Brushless micromotor is stuck or broken.
H 71 Romexis user ID already in use. Check the ID, select another one, or allow
rebind in Romexis.
H 72 Incompatible Romexis version. PlanID is not supported with this version of
Romexis.
H 72.1 Incompatible Romexis version User settings for currently active instrument
can not be used. Romexis version does not
support user instrument settings or presets
for currently active instrument. Check the
instrument settings before using instrument
and update Romexis.
H 73 Failed to copy user profile from Romexis.
H 73.1 Failed to save user settings.
H 74 PlanID reader missing. PlanID reader is either missing or not
functional.
H 75 Yearly maintenance in <#> days. Contact service for yearly maintenance.
H 76 Yearly maintenance performed.
H 79.1 Serial number chip missing. The serial number chip is missing. Contact
service.
H 80 Headrest length safety feature activated
when driving headrest. Check that nothing is
blocking headrest movements.
Check that nothing is blocking the headrest
movement. The headrest can be driven
normally after the possible obstruction has
been removed.
H80.1 Headrest length drive limit activated when
driving headrest.
Check that nothing is blocking headrest
movements.
Faulty length motor, motor cable,
nonlubricated slider rail or position
potentiometer may cause this problem.
27 Help and error messages
146 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Classic v2 and is the answer not in the manual?

Planmeca Compact i Classic v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Classic v2
CategoryDental equipment
LanguageEnglish

Related product manuals