Factory presets for image resolution Normal in programs Nose and Sinus
Patient size kV value mA value mA value with
ULD
Child (XS) 100 3.6 1.8
Small adult (S) 120 3.2 1.6
Medium-sized adult (M) 120 4 2
Large adult (L) 120 5 2.5
Extra large adult (XL) 120 6.3 3.2
Factory presets for image resolution Normal in other programs
Patient size kV value mA value mA value with
ULD
X-ray units with 90 kV X-ray tube
Child (XS) 90 5.6 3.6
Small adult (S) 90 6.3 4.5
Medium-sized adult (M) 90 8 5.6
Large adult (L) 90 10 7.1
Extra large adult (XL) 90 12.5 9
X-ray units with 120 kV X-ray tube (D-059SBR)
Child (XS) 100 6.3 2.8
Small adult (S) 120 5 2.5
Medium-sized adult (M) 120 6.3 3.2
Large adult (L) 120 8 4
Extra large adult (XL) 120 8 5
X-ray units with 120 kV X-ray tube (SXR 130-10-0.5 SC)
Child (XS) 100 6.3 2.8
Small adult (S) 120 5 2.5
Medium-sized adult (M) 120 6.3 3.2
Large adult (L) 120 8 4
Extra large adult (XL) 120 8 5
Factory presets for image resolution High Definition (HD)
Patient size kV value mA value mA value with
ULD
X-ray units with 90 kV X-ray tube
Child (XS) 90 6.3 4.5
Small adult (S) 90 8 5.6
Medium-sized adult (M) 90 10 7.1
Large adult (L) 90 12.5 9
Extra large adult (XL) 90 14 11
X-ray units with 120 kV X-ray tube (D-059SBR)
8 3D patient exposure
User's manual Planmeca ProMax 39