15 Service
The X-ray unit must be checked by a qualified Planmeca service technician
once a year or after every 10 000 exposures (if this is sooner). This will
guarantee patient and user safety and ensure consistent image quality.
The annual maintenance service includes inspection of the following:
X-ray unit adjustments and quality control checks
Data security updates
Exposure switch
Exposure indicator lights and warning signals
Emergency stop button
Column motor nut
Labels