Page 32 of 32 Introduction
Table 1-1 (continued)
Front Panel Controls and Indicators
Fig 1-1
Ref. # Item Function
9 UTILITY Pressing this key will allow the user to set functional changes which affect
KEY all tests. These functions are listed below and discussed in detail in the
Operation Section of this manual.
Measure Offset- Off, On, Press Start - Get Offset
Keypad Lockout- Off, Lock Password,
Lock Cycle Tests Password
Test Locations- Show All Tests, Hide Disabled Tests
Buzzer on Fail- Off, Soft, Loud
Beep on Key Hit- On, Off
IEEE Address- 1 thru 16
Enhanced Features- On, Off
Test Low Limit- Continuous, End of Test
Delay Time- Off, 0.1 - 999.9sec
Cont Volt On Fail- Off, On
Cont Steps On Fail- Off, On
Stop Test On Pass- Off, On
Step And Increment- Off, On
Clear All Tests- No, Yes
Test To View- All Tests, Only Sequence
Only Single Step, Only Factory
Perform Calibration- Cal Date, Start Full, Start GC Only
Ground Fault Stop- Off, On, Off - DUT Grounded
Arc Pulse Width- 5 - 500µsec
Pressing UTILITY key at any time exits the utility mode back to Idle.
10 LEFT/RIGHT In Idle mode moves to the next or previous parameter for viewing.
Arrow Keys In Program mode moves to the next or previous parameter for
programming.
In Utility mode moves to the next utility function.
11 UP/DOWN In Idle mode moves to the next or previous test for viewing.
Arrow Keys In Program mode changes the condition of the selected test parameter.
In Utility mode changes the condition of the selected utility function.
12 LOCKOUT When illuminated indicates that the front panel keypad lockout has been
Indicator enabled. (Illuminates green on late model instruments)