Interface Page 83 of 83
Table 3-4: IEEE & RS-232 Commands (continued)
Command Function Parameter(s)
Example: Sequence 1 reply with no calc tests programmed
User sends VIEW:TEST?:SEQ 1 and response is
S1:1 Test: nn (or none)
S1:2 Test: nn (or skip, or none)
S1:3 Test: none
Example: Sequence 2 reply with step 3 programmed for calc
User sends VIEW:TEST?:SEQ 2 and response is
S2:1 Test: 24
S2:2 Test: 25
S2:3 Test: calc
S2:3 Calc: AC Real Current
S2:3 Calc Hi Lim: 1.000mA
S2:3 Calc Low Lim: 0.000mA
UTILity? View the utilities programmed
If enhanced features is set to On the format of the returned data is as follows:
Meas Offset:<tab><Off, On, or Get><CR>
Keypad Lockout:<tab><Off, On, or On With Test Number
><CR>
Test Locations:<tab><Hide Disabled Tests or Show All
Tests><CR>
Buzzer on Fail:<tab><Loud, Soft, or Off><CR>
Beep on Key Hit:<tab><Off or On><CR>
RS-232 Baud Rate:<tab><9600 or 19200><CR>
IEEE Address:<tab><nn><CR>
Test Low Limit:<tab><Continuous or End Of Test><CR>
Delay Time:<tab><nnn.nsec or Off><CR>
Cont Voltage On Fail:<tab><Off or On><CR>
Cont Steps On Fail:<tab><Off or On><CR>
Stop Test On Pass:<tab><Off or On><CR>
Step And Increment:<tab><Off or On><CR>
Tests To View:<tab><All Tests, Only Single Step, or
Only Sequence><CR>