ppbRAE 3000 User’s Guide
99
To install the adapter in the instrument:
1. Remove the Li-ion battery pack from the battery
compartment by sliding the tab and tilting out the battery.
2. Replace it with the alkaline battery adapter
3. Slide the tab back into place to secure the battery adapter.
IMPORTANT!
Alkaline batteries cannot be recharged. The instrument’s internal
circuit detects alkaline batteries and will not allow recharging. If you
place the instrument in its cradle, the alkaline battery will not be
recharged. The internal charging circuit is designed to prevent
damage to alkaline batteries and the charging circuit when alkaline
batteries are installed inside the instrument.
Note: When replacing alkaline batteries, dispose of old ones properly.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the
battery only in areas known to be non-hazardous. Remove and replace
the battery only in areas known to be non-hazardous.
Note: The internal charging circuit is designed to prevent charging to
alkaline batteries.