EB200 Manual SENSe subsystem FUNCtion
4052.2000.02 4.75 E-7
. :FUNCtion
If the sensor function(s) is (are) changed, the trace data set MTRACE is always deleted.
. . :CONCurrent <Boolean>
Determines whether several sensor functions can at the same time be switched or not. If CONCurrent =
OFF, the command SENSe:FUNCtion[:ON] has the effect of a 1-out-of-n selection (one is switched
on, the previously activated is automatically switched off). If CONCurrent = ON, the command
SENSe:FUNCtion[:ON] switches the corresponding function on, while all the other functions remain
unchanged. If CONCurrent is switched from ON to OFF, the function "VOLTage:AC" is switched on
and all other functions are switched off.
Parameters:
ON CONCurrent on
OFF CONCurrent off
*RST state:
ON
Example:
FUNCtion:CONCurrent ON
. . :CONCurrent?
Query of several sensor functions that, at the same time, can be switched or not.
Parameters:
none
Result:
CONCurrent switched on 1
CONCurrent switched off 0
Example:
FUNCtion:CONCurrent? -> 1
. . :OFF <sensor_function> {,<sensor_function>}
Switch off of one or several sensor functions.
Parameters:
see SENSe:FUNCtion[:ON]
*RST state:
"FREQ:OFFS"
Example:
FUNCtion:OFF "FREQ:OFFS"