EB200 Manual SENSe subsystem FUNCtion
4052.2000.02 4.76 E-7
. . :OFF?
Query of the sensor functions being switched off
Parameters:
none
Result:
List of the sensor functions being switched off. For strings see SENSe:FUNCtion[:ON]
Example:
FUNCtion:OFF? -> "FREQ:OFFS"
. . . :COUNt?
Query of the number of sensor functions being inactive
Parameters:
none
Result:
Number of sensor functions being inactive
Example:
FUNCtion:OFF:COUNt? -> 2
. . [:ON] <sensor_function> {,<sensor_function>}
Switch on of one or several sensor functions.
Error message:
If CONCurrent = OFF, an error -108, "Parameter not allowed" will be generated for two or
several parameters.
Parameters:
<sensor_function> is one of the following strings:
"VOLTage:AC" switch on level measurement
"FREQuency:OFFSet" switch on offset measurement
"FSTRength" switch on field strength measurement
Note:
The sensor function FSTRength will only provide results if the SW option EB200FS is fitted. Also see
Annex H.
*RST state:
"VOLTage:AC"
Example:
FUNCtion "VOLT:AC", "FREQ:OFFS"