EB200 Manual SENSe subsystem FUNCtion
4052.2000.02 4.77 E-7
. . [:ON]?
Query of sensor functions being switched on
Parameters:
none
Result:
List of sensor functions switched on. If no function is active, a zero string ("") is output. The list has a
specific order:
1. level measurement function
2. offset measurement function
3. field strength measurement function
The following strings are to be expected:
"VOLT:AC" level measurement switched on
"FREQ:OFFS" offset measurement switched on
"FSTR" field strength measurement switched on
Example:
FUNCtion? -> "VOLT:AC", "FREQ:OFFS"
. . . :COUNt?
Query of the number of sensor functions being active
Parameters:
none
Result:
Number of sensor functions being active
Example:
FUNCtion:Count? -> 2