EB200 Manual SENSe subsystem MSCan
4052.2000.02 4.83 E-7
. . :DIRection?
Query of scan direction
Parameters:
none
Result:
UP, DOWN
Example:
MSCan:DIRection?
DOWN
. . :DWELl <numeric_value>|MINimum|MAXimum|INFinity
Setting the dwell time with the hold criterion fulfilled (T-DWELL in MSCAN-CONFIG menu).
Note:
According to the SCPI standard, this command is used to set the dwell time per scan step, ie the time
required by a step. This definition is met in EB200, if the squelch is switched off. The hold criterion is
then fulfilled for each step.
Parameters:
<numeric_value> dwell time in seconds
MINimum|MAXimum lowest|highest dwell time
INFinity infinite dwell time
*RST state:
0.5 s
Example:
SWEep:DWELl 10 ms
. . :DWELl? [MINimum|MAXimum]
Query of dwell time
Parameters:
none query of current dwell time
MINimum|MAXimum query of lowest|highest dwell time
Result:
Dwell time in seconds; 9.9E37 is output for an infinite number
Example:
MSCan:DWELl? 0.010