EasyManua.ls Logo

R&S EB200 - Page 158

R&S EB200
232 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EB200 Manual SENSe subsystem MSCan
4052.2000.02 4.84 E-7
. . :HOLD
. . . :TIME <numeric_value>|MINimum|MAXimum
Setting the hold time during signal-controlled scan continuation (T_NOSIG in MSCAN-CONFIG menu).
If the signal disappears during the dwell time, the hold time is started. After completion of the hold time,
the scan is continued with the next frequency even if the dwell time has not yet been completed. If the
signal during the hold time exceeds the squelch threshold, the hold time is reset and the end of the
dwell time or the renewed disappearance of the signal is awaited. The hold time is only important if the
control function "STOP:SIGNal" (see SENSe:MSCan:CONTrol) is switched on.
Parameters:
<numeric_value> hold time in seconds
MINimum|MAXimum lowest|highest hold time
*RST state:
0 s
Example:
SWEep:HOLD:TIME 10 ms
. . . :TIME? [MINimum|MAXimum]
Query of hold time
Parameters:
none query of current hold time
MINimum|MAXimum query of lowest|highest hold time
Result:
Hold time in seconds
Example:
MSCan:HOLD:TIME? 0.010

Table of Contents

Related product manuals