EB200 Manual SENSe subsystem LIST
4052.2000.02 4.85 E-7
. :LIST
Within EB200 the LIST subsystem is used to define a list of frequencies. Provided that the EB200CM
SW option is fitted, in
FREQuency:MODE LIST mode triggered level measurements can be carried
out on up to 50 frequencies.
. .
:FREQuency <numeric_value>, <numeric_value>, ....
Input of list of frequencies on which levels are to be measured in LIST mode.
Parameter:
<numeric_value> frequency value
Note:
The number of possible frequency values can be queried by using command
SENSe:LIST:FREQuency:POINts?.
*RST state:
935200000, 935400000, 935600000
50 frequencies altogether in increments of 10 kHz
Example:
LIST:FREQuency 935.8 MHz, 936.0 MHz, 936.2 MHz
. . LIST:FREQuency?
Query of list of frequencies on which levels are to be measured in LIST mode
Parameters:
none
Result:
<numeric_value>, <numeric_value>, ...
Example:
LIST:FREQuency? -> 935800000, 936000000, 936200000
. . LIST:FREQuency:POINts? MINimum|MAXimum
Query of the number of frequencies covered by command
LIST:FREQuency
Parameters:
none query of current number of frequencies
MINimum|MAXimum query of minimum/maximum number of frequencies
Example:
LIST:FREQuency:POINts? MAX -> 50