EB200 Manual SENSe subsystem SWEep
4052.2000.02 4.93 E-7
. . :SUPPress
Insert current frequency into suppress list. The range is obtained from the bandwidth according to the
following formulae:
SSTARTn = SENSn: FREQ - SENSn:BAND/2
SSTOPn = SENSn: FREQ + SENSn:BAND/2
The frequency pair is inserted into an empty space of the trace. Free spaces (gaps) are characterized
by a frequency pair with the values 0.0.
Error message:
If the corresponding suppress trace has no free space, an error -223 "Too much data" is
generated.
Note:
Suppress list 1 to 100 corresponds to 0 to 99 on front panel
Parameters:
none
*RST state:
none, as command is an event
Example:
SWEep:SUPPress
. . . :SORT
Sort and condense suppress list. The frequency pairs are sorted to an ascending order of the start
frequency. Overlapping is eliminated by extending the frequency pair. The other frequency pair is then
deleted. Gaps within the suppress list are put to the end of the list.
Parameters:
none
*RST state:
none, as command is an event
Example:
SWEep:SUPPress:SORT