EB200 Manual TRACe subsystem
4052.2000.02 4.117 E-7
. :POINts? <trace_name>[,MINimum|MAXimum]
Query of number of values stored in a trace
The number of values stored in the suppress traces is always 100. Thus, the MAXimum and MINimum
value is also 100.
The number of values stored under IFPAN depends on the IF bandwidth and is between 770 and 1230.
Error message:
If the trace name is unknown, an error -141, "Invalid character data" will be generated.
Parameters:
<trace_name> see TRACe[:DATA]?
no further parameter query of current number
MINimum|MAXimum query of lowest|highest number
Result:
Number of values
Example:
TRACe:POINts? MTRACE, MAX -> 2048
. . :AUTO? <trace_name>
Query of trace length (auto-adjust)
A 0 (no auto-adjust for trace length) is always output for a suppress trace.
Error message:
If the trace name is unknown, an error -141, "Invalid character data" will be generated.
Parameters:
<trace_name> see TRACe[:DATA]?
Result:
0 no auto-adjust for trace length
1 auto-adjust for trace length
Example:
TRACe:POINts:AUTO? MTRACE;AUTO? ITRACE -> 1;1