EB200 Manual Remote control
4052.2000.02 4.17 E-7
*RST
RESET sets the device to a defined default status. The default setting is indicated in the description
of the commands.
*SRE 0 to 255
SERVICE REQUEST ENABLE sets the service request enable register to the value indicated. Bit 6
(MSS mask bit) remains 0. This command determines under which conditions a service request is
triggered. Query *SRE? reads the contents of the service request enable register in decimal form.
Bit 6 is always 0.
*STB?
READ STATUS BYTE QUERY reads out the contents of the status byte in decimal form.
*TRG
TRIGGER triggers the same actions as the INITiate:CONM[:IMMediate] command.
*TST?
SELFTEST QUERY triggers the module state test and yields a figure which is to be interpreted as
the bit field:
Result = 0 -> all modules ok
Result
≠ 0 -> fault in one or several modules. The information about the possible error can be
queried by means of the SYSTem:ERRor? Command.
*WAI
WAIT-to-CONTINUE only permits the servicing of the subsequent commands after all preceding
commands have been executed and all signals have settled (also see "Status reporting system" on
page 127 and "*OPC").