EB200 Manual CALCulate subsystem
4052.2000.02 4.22 E-7
. :DSCan
. . :AVERage
. . . :TYPE MAXimum|OFF
Setting of the display mode of the DSCan spectrum
Parameters:
MAXimum indication of maximum values
OFF indication of current values
*RST state:
OFF
Example:
CALCulate:DSCan:AVERage:TYPE MAXimum
. . . :TYPE?
Query of the display mode of the DSCan spectrum
Result:
MAX indication of maximum values
OFF indication of current values
Example:
CALCulate:DSCan:AVERage:TYPE? -> MAX
. . :CLEar
Clears the DSCan spectrum. If the display has been selected to be
MAXimum, maximum value
generation restarts.
Parameters:
none
Example:
CALCulate:DSCan:CLEar