3.3 SERVICE TOOLS
S8 Test Filter Wheel
The rotation of the filter wheel can be controlled.
With [F0] up to [F9] the discrete positions of the filter
wheel can be selected.
A specific value (e.g. 399MS) shows the rotation of the
filter wheel. The abbreviation MS is used for motor
step.
TEST FILTER WHEEL
F1 340nm 399MS
F1 F2 F3 F4 F5 F6 F7 F8
ESCF9 F0
S9 Temperature ON/OFF
The temperature regulation can be switched on or
switched off. The current setting is shown in the first
line as ON or OFF.
The display shows additionally the following
information:
• a D in case of an inserted flow through cuvette
adaptor
• an E in case of an inserted standard cuvette
adaptor
• the required value of temperature regulation
• the actual value of temperature regulation
• the stability value
• the current value of ADC
With [SELECT] the temperature level can be chosen.
With [OK] the temperature regulation is started.
[EXIT] leaves the function without any changes.
TEMPERATURE 37C ON D
OFF REQUIRED 14165
25C ACTUAL 14053
30C STABILITY 4
37C CONVERTER 947
EXIT OK SELECT
RIELE BERLIN Photometer 5010 V5+ 3.3.8 11.03.2009