3.3 SERVICE TOOLS
S21 Clear factors + RB
The determined resulting factor of a standard
measurement and the reagent blank (RB) is stored
together with the corresponding method number. After
renewed selection of this method the last resulting
factor is offered as "OLD STD", the reagent blank (RB)
is offered as “OLD RB”.
All
resulting factors and reagent blanks (RB) will be
erased by invoking [START].
[EXIT] leaves the function without any changes.
CLEAR FACTORS + RB
EXIT START
S22 Clear error list
The last 10 serious errors are shown or printed.
The complete error list will be erased by invoking
[START].
[EXIT] leaves the function without any changes.
CLEAR ERROR LIST
EXIT START
S23 Printer test
A special printout for testing the printer ribbon or the
hardware is produced.
The printout is started by invoking [START].
[EXIT] leaves the function without any changes.
PRINTER TEST
EXIT START
RIELE BERLIN Photometer 5010 V5+ 3.3.14 11.03.2009