4.20 BUBBLE DETECTOR WITH PLUG REF 1706870001
Function Description
The bubble detector controls the aspirated volume by
an integrated light barrier. The detector gives an
analogue signal showing the status AIR or WATER.
The signal is controlled by the microcontroller and
produces in case of a wrong handling or technical
defective an error message.
For further information see OPERATING
INSTRUCTIONS, chapter 9 – Error message /
Correction.
Location
Bubble detector
Replacement
Open the lid. Disconnect the bubble detector. Withdraw the aspiration tubing from the metal tube. Don’t bend
the apiration tubing. Replace the bubble detector.
Installation
For installation proceed in reverse order.
For further information see OPERATING INSTRUCTIONS, chapter 7.2.4 – Pump / bubble detector.
RIELE BERLIN Photometer 5010 V5+ 4.20.1 11.05.2006