3.3 SERVICE TOOLS
S10 Temperature Value
The temperature values of the device can be read and
set. A new value can be entered with maximum five
digits.
With [SELECT] a specific temperature value can be
set. The temperature values are marked as E25, E30,
E37, D25, D30 and D37 corresponding to the type of
adaptor (E or D) and the temperature level (25° C,
30° C and 37° C).
The cursor of the current temperature value is blinking.
With [NEW] a temperature value can be edited.
Enter temperature value. Confirm with [E].
(chapter 4.40 - Thermal adjustment)
TEMPERATURE VALUE
E25 8322 D25 8299
E30 9552 D30 9485
E37 11273 D37 11165
EXIT NEW SELECT
NEW VALUE
1 2 3 4 5 6 7 8
ESC 9 0 . + - E
S11 Correction Value
The correction values of the device can be read and
set. A new value can be entered with maximum five
digits.
With [SELECT] a specific correction value can be set.
The correction values are marked as E25, E30, E37,
D25, D30 and D37 corresponding to the type of
adapter (E or D) and the temperature level (25° C,
30° C and 37° C).
The cursor of the current correction value is blinking.
With [NEW] a correction value can be edited.
Enter correction value. Confirm with [E].
CORRECTION VALUE
E25 3000 D25 3000
E30 3000 D30 3000
E37 3000 D37 3000
EXIT NEW SELECT
NEW VALUE
1 2 3 4 5 6 7 8
ESC 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.9 11.03.2009