4.31 TOP CASE PRE-ASSEMBLED REF 5010-084
WITH TOUCHSCREEN
RIELE BERLIN Photometer 5010 V5+ 4.31.1 15.04.2009
Function Description
The top case is one of the two main parts of the
housing. It is assembled with:
Sipping lever
Microswitch (see chapter 4.33)
Metal tube inlet
Lid
Touchscreen with PCB RR-122
(see chapter 4.41)
Plastic foil touchscreen
(see chapter 4.34)
Location
Top case
Replacement
The Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening – item 1, item 2 and item 5
see chapter 5.2 Separating the top case from the bottom case
see chapter 4.41 Touchscreen
see chapter 4.43 PCB RR-122 Display, printer, touch
see chapter 4.34 Plastic foil touchscreen
Note: When the plastic foil touchscreen (REF 5010-070) is removed from device it might not be reusable.
Installation
For installation proceed in reverse order.