3.3 SERVICE TOOLS
S15 ADC Counts (Dark Adjustment)
The ADC values of the dark adjustment can be read.
The display shows the following information:
• a D in case of a preselected flow through cuvette
adaptor
• an E in case of a preselected standard cuvette
adaptor
• the ADC values assigned to the amplification levels
V0 up to V7
With [E/D] the display shows alternately the values in
combination with the different adaptors E or D.
ADC COUNTS (DARK ADJ.)
E
0: 18425 1: 18428
2: 18433 3: 18442
4: 18457 5: 18483
6: 18534 7: 18630
EXIT E/D
S16 Boost Wavelength
To improve the low light gain at small wavelengths, the
measurement time is extended when operating below a
definite wavelength. This definite wavelength will be
displayed on the right upper corner of the window.
The default value is 390 nm.
The input has to be confirmed with [E].
BOOST WAVELENG 390nm
1 2 3 4 5 6 7 8
ESC 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.11 11.03.2009