RIGOL Chapter 2 Command System
2-190 MSO1000Z/DS1000Z Programming Guide
:TRIGger:WINDows:POSition
Syntax
:TRIGger:WINDows:POSition <pos>
:TRIGger:WINDows:POSition?
Description
Set or query the trigger position in windows trigger.
Parameter
Name
Type
Range
Default
<type>
Discrete
{EXIT|ENTER|TIMe}
ENTER
Explanation
EXIT: trigger when the input signal exits the specified trigger level range.
ENTER: trigger when the trigger signal enters the specified trigger level range.
TIMe: used to specify the hold time of the input signal after it enters the specified
trigger level range. The instrument triggers when the accumulated hold time equals
the windows time.
Return
Format
The query returns EXIT, ENTER, or TIM.
Example
:TRIGger:WINDows:POSition ENTER /*Set the trigger position to enter*/
:TRIGger:WINDows:POSition? /*The query returns ENTER*/
:TRIGger:WINDows:TIMe
Syntax
:TRIGger:WINDows:TIMe <NR3>
:TRIGger:WINDows:TIMe?
Description
Set or query the hold time in windows trigger.
Parameter
Name
Type
Range
Default
<NR3>
Real
8ns to 10s
1μs
Return
Format
The query returns the hold time in scientific notation.
Example
:TRIGger:WINDows:TIMe 0.002 /*Set the hold time to 2ms*/
:TRIGger:WINDows:TIMe? /*The query returns 2.000000e-03*/
Related
Command
:TRIGger:WINDows:POSition

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope
Input Impedance1 MΩ ± 2%