Appendix • A
165
Operator ID
Operator ID entry control 0: None
1: Prompt
2: Scan only
3: Prompt (numeric only)
1YY
Operator ID entry control on Glucose con-
trol only
0: no (always)
1: yes (on controls only)
0YY
Operator ID validation
(Allowed characters: a-z, 0-9,
“.” [period], “-” [hyphen])
0: none
1: length
2: list
3: list & password
0YY
Operator ID maximum length (used to
validate an operator ID if operator ID
validation mode is set to “Length”)
0 – 20 20 Y Y
Operator ID minimum length (used to
validate an operator ID if operator ID
validation mode is set to “Length”)
0 – 20 0 Y Y
Operator ID timeout (determines the time in
sec that an operator remains logged in after
power-off). 0 = feature disabled.
0 – 3,600sec 0 N Y
Operator password length 4-20 characters (only a-z,
0-9 allowed)
N/A N N
Operator certification expiry warning 0-90 days 0 (=off) N Y
Operator Name length 0-25 characters N/A N N
Operator ID and patient ID barcode masks
Operator ID barcode mask (see separate
table at the end of this chapter)
0 – 300 characters* “” N Y
Patient ID barcode mask (see separate
table at the end of this chapter)
0 – 300 characters* “” N Y
* The number of characters in the unmasked barcode may not exceed 20.
Subject/Attribute Range Default Device DMS