Quality Control
113
The first steps in this quality control test are identical to the procedure described before, see
page 106. The following description therefore starts at the main menu.
3 Touch QC Test.
4 Touch in the QC Status screen to con-
tinue with quality control test. The test
strip icon prompts you now to insert a test
strip.
5 Remove the test strip from its IQC test
strip container.
6 Insert the test strip as described before,
see page 107.
7 If you are using a new lot of IQC test strips
and have not inserted the code chip yet,
you must do so now. If you have used the
test strips already, skip this step.
Main Menu
11/24/2009
09:15 am
QC Test
Memory
Setup
Patient TestQC!
QC Test - QC Status
09:15 am
CK-MB
QC Status
Myo
DDimer
- Pass
- Pass
- Pass
Trop T
proBNP
- Locked
- Locked
QC Test - Ins. Strip
09:15 am
Insert Strip
JONES, TOMOp:
QC Test - Ins. Chip
IQC
09:15 am
LOT 226 437-02
JONES, TOM
Par:
QC
Op:
Insert Code
Chip
4529