Performing a Test
92
Performing a Test
1 Check the charge level of the handheld
battery pack. If there are no bars left in the
battery icon, you cannot perform any
more tests.
2 Check that the date and time are correct.
Correct any incorrect settings as
described in the chapter entitled “Meter
Setup/Setting the Date”.
If a QC lockout is displayed instead of Patient
Test, you must run a quality control test
before you can perform a test (see “Quality
Control” starting on page 105). When aQC
Lockout is in force, a test can only be per-
formed only as a STAT test.
If a Cleaning Lockout is displayed instead of
Patient Test, you must touch Cleaning and
follow the instructions on the screen and
clean (if required) the meter. See “Cleaning”
beginning on page 123. Once you have
replaced the sample application cover, the
lockout is no longer in force once you switch
on the meter again.
If the Patient Test button is available, but a
lockout is displayed, you must first perform a
quality control test for certain test parameters.
Other test parameters are not affected.
Note for international users: Due to limited
space on the display some screens may show
“QC” instead of the localised version of this
term.
Main Menu
11/24/2009
09:15 am
QC Test
Memory
Setup
Patient Test
Logout
Main Menu
11/24/2009
09:15 am
QC Test
Memory
Setup
Lockout
Logout
Main Menu
Cleaning Lockout
09:15 am
Cleaning
Memory
Setup
Clean Meter
09:15 am
Remove Cover
Main Menu
11/24/2009
09:15 am
QC Test
Memory
Setup
Patient Test
Logout
QC!