Cleaning
124
Basic sequence of steps for cleaning
First remove any blood and other dirt using water or soapy water.
Then disinfect the meter.
Cleaning after Contamination due to Mispipetting
For detailed cleaning instructions (“how to clean…”) for the individual components of the meter,
see the following pages. The instructions below describe only the sequence of steps to be taken
in case of mispipetting.
Remove the sample application cover.
Remove and dispose of the test strip.
Clean the soiled parts of the meter with a moistened cotton bud or tissue.
Disinfect the meter.
NOTICE
Do not move the meter to prevent liquids from running any further.
NOTICE
Dab away all visible blood (also on the test strip, but not in the
application area).
NOTICE
Do not use the meter before the liquid/moisture has been completely
removed. During the drying time, the meter must not be moved as this
may soil the optical components.