EasyManuals Logo

Roche cobas h 232 User Manual

Roche cobas h 232
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
Cleaning
124
Basic sequence of steps for cleaning
First remove any blood and other dirt using water or soapy water.
Then disinfect the meter.
Cleaning after Contamination due to Mispipetting
For detailed cleaning instructions (“how to clean…”) for the individual components of the meter,
see the following pages. The instructions below describe only the sequence of steps to be taken
in case of mispipetting.
Remove the sample application cover.
Remove and dispose of the test strip.
Clean the soiled parts of the meter with a moistened cotton bud or tissue.
Disinfect the meter.
NOTICE
Do not move the meter to prevent liquids from running any further.
NOTICE
Dab away all visible blood (also on the test strip, but not in the
application area).
NOTICE
Do not use the meter before the liquid/moisture has been completely
removed. During the drying time, the meter must not be moved as this
may soil the optical components.

Table of Contents

Other manuals for Roche cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas h 232 and is the answer not in the manual?

Roche cobas h 232 Specifications

General IconGeneral
BrandRoche
Modelcobas h 232
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals