Performing a Test
83
Performing a Test
What you need:
■ Your cobas h 232 meter
■ Roche CARDIAC test strips for the desired test, with the supplied code chip
■ Roche CARDIAC pipettes (or suitable pipettes with pipetting volume of 150 µL)
Always …
■ … operate the meter according to the specified operating conditions (see page 135).
■ … follow the information on correct handling of test strips in the package insert.
■ … place the meter on a level, vibration-free surface while applying the sample until the
necessary sample has been absorbed completely by the test strip.
■ … keep the meter clean. See “Cleaning” on page 123.
Never …
■ … touch or remove the test strip during a test.
■ … add more blood after the test has begun.
■ … remove or insert the code chip while the meter is performing a test.
■ … move or pick up the meter while applying the sample - wait until the necessary sample
has been absorbed completely by the test strip.
■ … store the meter at extreme temperatures.
■ … store the meter in damp or humid conditions without protection.
Note: The Roche CARDIAC family includes several meters. Always make sure that the dis-
posables (e.g. test strips) are made for use with the cobas h 232 meter. Some assays might
require a minimum software version, please ensure that your software version on the
cobas h 232 fits to your assay as mentioned in the corresponding package insert. Please
contact your Roche representative if the cobas h 232 meter requires a software update
CAUTION
Accuracy/precision of measured results:
Failure to comply with the above may lead to inaccurate results. An
incorrect result may lead to an error in diagnosis, therefore posing
danger to the patient.