Meter Setup
78
STAT Test Configuration
Quality control tests ensure consistent quality and accurate measurements. In emergency situa-
tions, however, it may be necessary to perform a test without delay. To override an active lockout
for a test parameter, lot or operator, you can allow for STAT Tests (STAT = Short TurnAround
Time).
You can set a number of tests permitted beyond the lockout. Once the number of allowed STAT
Tests has been reached (for a single parameter), additional tests for this parameter are blocked
until you successfully perform a quality control test.
STAT tests are labelled accordingly in the result window.
Note: STAT Tests will be counted separately for each test parameter, lot or operator. Thus the
actual availability of STAT Tests may differ for each parameter.
Note for international users: Due to limited space on the display some screens may show
“STAT” instead of the localised version of this term.